TY - JOUR
T1 - Comparison of four rapid diagnostic tests, ELISA, microscopy and PCR for the detection of Giardia lamblia, Cryptosporidium spp. and Entamoeba histolytica in feces
AU - Van den Bossche, Dorien
AU - Cnops, Lieselotte
AU - Verschueren, Jacob
AU - Van Esbroeck, Marjan
N1 - NPP
PY - 2015
Y1 - 2015
N2 - PURPOSE: Microscopy is the diagnostic reference standard for the detection of parasites, but it is labor-intensive and requires experience. Rapid diagnostic tests (RDTs) can provide an alternative to microscopy.METHODS: RDTs from four different manufacturers were compared to enzyme-linked immunosorbent assay (ELISA), microscopy and/or parasite-specific real-time PCR: ImmunoCardSTAT!®CGE (Meridian Bioscience Inc., Cincinnati, Ohio, USA) (A), Crypto/Giardia Duo-Strip (Coris Bioconcepts, Gembloux, Belgium) (B), RIDA®QUICK Cryptosporidium/Giardia/Entamoeba Combi (R-BioPharm, Darmstadt, Germany) (C) and Giardia/Cryptosporidium Quik Chek (Techlab Inc., Blacksburg, Virginia, USA) (D).RESULTS: Thirty frozen samples were analyzed retrospectively. For Giardia lamblia (n=12) and Cryptosporidium (n=12) sensitivities ranged from 58% (B), over 83% (A, C) to 100% (D) and from 92% (B) to 100% (A, C, D), respectively. Specificity for both G. lamblia and Cryptosporidium was 100% for all RDT brands. Sensitivity for Entamoeba histolytica (n=5) was 100%, while specificity reached 80% (A) to 88% (C). In a prospective study, fresh samples were tested. For G. lamblia (n=30), sensitivity ranged from 66% (B), over 79% (A) and 83% (C) to 100% (D) and specificity varied between 94% (D) and 100% (A, B, C). For Cryptosporidium (n=3), sensitivity was 100% for all brands except (B) (67%) and specificities were 95% (A, B), 98% (C) and 100% (D). E. histolytica (n=1) was detected by both (A) and (C), while specificity was 81% and 87% respectively.CONCLUSION: RDTs can be a valuable tool when microscopic expertise is poor and in remote and outbreak settings where other techniques are often not available and rapid diagnosis is required.
AB - PURPOSE: Microscopy is the diagnostic reference standard for the detection of parasites, but it is labor-intensive and requires experience. Rapid diagnostic tests (RDTs) can provide an alternative to microscopy.METHODS: RDTs from four different manufacturers were compared to enzyme-linked immunosorbent assay (ELISA), microscopy and/or parasite-specific real-time PCR: ImmunoCardSTAT!®CGE (Meridian Bioscience Inc., Cincinnati, Ohio, USA) (A), Crypto/Giardia Duo-Strip (Coris Bioconcepts, Gembloux, Belgium) (B), RIDA®QUICK Cryptosporidium/Giardia/Entamoeba Combi (R-BioPharm, Darmstadt, Germany) (C) and Giardia/Cryptosporidium Quik Chek (Techlab Inc., Blacksburg, Virginia, USA) (D).RESULTS: Thirty frozen samples were analyzed retrospectively. For Giardia lamblia (n=12) and Cryptosporidium (n=12) sensitivities ranged from 58% (B), over 83% (A, C) to 100% (D) and from 92% (B) to 100% (A, C, D), respectively. Specificity for both G. lamblia and Cryptosporidium was 100% for all RDT brands. Sensitivity for Entamoeba histolytica (n=5) was 100%, while specificity reached 80% (A) to 88% (C). In a prospective study, fresh samples were tested. For G. lamblia (n=30), sensitivity ranged from 66% (B), over 79% (A) and 83% (C) to 100% (D) and specificity varied between 94% (D) and 100% (A, B, C). For Cryptosporidium (n=3), sensitivity was 100% for all brands except (B) (67%) and specificities were 95% (A, B), 98% (C) and 100% (D). E. histolytica (n=1) was detected by both (A) and (C), while specificity was 81% and 87% respectively.CONCLUSION: RDTs can be a valuable tool when microscopic expertise is poor and in remote and outbreak settings where other techniques are often not available and rapid diagnosis is required.
U2 - 10.1016/j.mimet.2015.01.016
DO - 10.1016/j.mimet.2015.01.016
M3 - A1: Web of Science-article
C2 - 25615719
SN - 0167-7012
VL - 110
SP - 78
EP - 84
JO - Journal of Microbiological Methods
JF - Journal of Microbiological Methods
ER -